Cross-reference

Replacing a Nafion™-based dryer or humidifier tube?

Find the nearest sunsep™ format by series and size, then send us your conditions. We'll recommend a configuration to evaluate in your instrument, including custom lengths and fittings when a standard part isn't an exact fit.

sunsep dryers and Aquadrive tubes

Nearest-format starting points

If you use (series) Format sunsep™ starting point Notes
MD-070 / MD-110 series12 to 144 in Single-tube sample dryer SWG-A01-03 to -36, 12 to 144 in Lengths match one to one (12, 24, 48, 72, 96, 144 in) and so do the ¼ in connections. Polypropylene (P) housings map to /PP and fluorocarbon (F) to /KF (PVDF). Up to 4 L/min, the same maximum Perma Pure publishes for the MD-110.MD-110 part numbers →
PD-50T series12 to 72 in Multi-tube sample dryer SWG-03524 and 48 in Perma Pure recommends 2 to 10 L/min for the PD-50T; SWG-035 runs up to about 20 L/min on AGC's curves. Ports: ¼ NPT supply, ⅛ NPT purge. SWG-035 is not offered in 12 or 72 in.PD part numbers →
PD-100T series Multi-tube sample dryer SWG-10012, 24 and 48 in Perma Pure recommends 4 to 20 L/min for the PD-100T; SWG-100 runs up to about 30 L/min on AGC's curves. Polypropylene or SUS316.PD part numbers →
PD-200T series12 to 72 in Multi-tube sample dryer SWG-10012, 24 and 48 in Perma Pure recommends 10 to 40 L/min for the PD-200T; SWG-100 runs up to about 30 L/min, so check your flow. Polypropylene or SUS316. No standard 72 in length.PD part numbers →
ME / BE-050 series Small-diameter micro-flow tube SWT-1.3 About 300 mL/min. Standard lengths 150, 300 and 600 mm, plus 1220 and 1830 mm with straight 4 mm ends; other lengths available. Ends: Luer (male/male or female/female), barb, straight 4 mm tube or stainless tube. MOQ 10.ME-050 part numbers →
ME / BE-060, 070, 110 series Larger micro-flow tube SWT-3.0 About 2 L/min, against 600 mL/min (ME-070) and 1 L/min (ME-110) published by Perma Pure. Ends: barb or straight 6 mm tube. Standard lengths 150, 300 and 600 mm, plus 1220 and 1830 mm. Custom-length assemblies available. MOQ 5. For BE-series parts, check your flow.ME-110 / ME-070 part numbers →

Matches are by format, size and connection, not a claim of identical performance. Membrane materials differ, and results depend on flow, pressure, temperature and purge. Confirm in your application before switching. sunsep™ figures come from the AGC sunsep™ catalog (2025); third-party figures come from published datasheets and distributor guides and may change. Third-party series names are used for identification only.

Bare membrane tubing is not a standard Brink product. If you are building a custom module, tell us what you need.

Medical or veterinary device?

We can work with medical, veterinary and life-science device makers, and the process is different. These dryers are industrial components that have not been evaluated for medical use. Evaluation samples are for non-clinical evaluation only and may not be used in clinical trials. Production use requires a medical-use agreement with the manufacturer, and the device maker takes responsibility for qualification, risk and regulatory approval. Tell us early and we'll walk you through it.

What stays the same

Both are unpowered, selective water-vapor membranes, usually run with a dry purge or sweep stream. Plumbing and purge setup are often similar.

What to check

Length, end fittings and bend radius in your enclosure; purge source and ratio; inlet temperature and condensation risk; pressure drop at your flow.

Why a second source

Many instrument makers qualify an alternative membrane part to reduce single-source supply risk. We support side-by-side evaluation.

Don't see your part?

Send the model number from the label or your bill of materials, plus flow, pressure, temperature and dew-point target. We reply within one business day with a recommended configuration.

Send my part number